DISPLAY ALARM INDICATORS 30
CAL Flashes if unit needs to be re-calibrated due to a 36F (20C) tempera-
ture change since last calibration. CAL is on constantly when unit is
outside of the -22°F to +158°F (-30C to +70C) operating temperature
range. CAUTION: Vertical measurements may be in error.
--0-- Periodic ash on the display indicates re-zero may be necessary
due to a temperature change of 18F (10C) since the last zero.
8888 Alternates with displayed elevation when the Measurement Module
is more than 20’ (6m) above or below the Base Unit. CAUTION:
Overrange measurements may be in error. If 8888 ashes well
within range, check troubleshooting chart; check for Cord damage
or leakage and if necessary, return for service.
BAT Flashes when only a few hours of battery life remain. A fresh 9V
alkaline battery can allow up to 60 hours of active use.
CORD Flashes to indicate need for routine factory Cord recharge required
about every 3 years or possible Cord damage. CAUTION: Measure-
ments may be in error.